From: Consistency and variability in human performance during simulate infant CPR: a reliability study
 | Mean (SD) | ICC (95% CI) | SEM | MDC | MDC as % of mean |
---|---|---|---|---|---|
Rate QI (%) | 38 (29) | .90 (.81–.95) | 9.26 | 25.67 | 67% |
Depth QI (%) | 45 (19) | .91 (.83–.96) | 5.64 | 15.63 | 35% |
Leaning QI (%) | 42 (21) | .94 (.88–.97) | 5.23 | 14.49 | 34% |
Duty cycle QI (%) | 18 (17) | .87 (.76–.94) | 6.03 | 16.73 | 93% |